1.0 PURPOSE:
To define the HPLC column performs satisfactorily and gives reproducible results.
2.0 SCOPE:
This SOP is applicable to a procedure for Operation, Cleaning, Storage, Calibration And Regeneration of HPLC Column ….
3.0 REFERENCE:
NA
4.0 RESPONSIBILITY:
4.1. QC Officer/ QC Executive
5.0 ACCOUNTABILITY:
QC Manager
6.0 PROCEDURE:
6.1 COLUMN OPERATING AND CLEANING
- On receipt of the new column put the date of received on the box of the column.
- While using the column for the first time, flush the column with shipping solvent (the details available.
- With manufacturer’s COA) for at least 30 minutes with a flow rate of 0.5 ml/min.
- Change the column to mobile phase, which is to be used for the column calibration. Allow the system to stabilize and condition the column for 30 minutes.
- Carry out the column calibration as mentioned in point No. 6.2.
- After calibration of the column, flush the column with respective solvents .
- Store the column in appropriate storage solvent as mentioned.
- Remove the column, fit the end caps and store in the column box.
6.2 CALIBRATION PROCEDURE
- Calibrate the column using the procedure tabulated in Annexure-II. For new columns not mentioned in Annexure-II, follow the method of calibration as specified in the manufacturer’s COA.
- Connect the column to HPLC system.
- Flush the column with mobile phase as mentioned in the procedure for respective columns.
- Prepare the test solution in the mobile phase as recommended in the procedure.
- Note down the required details on the record of analysis.
- After the column is conditioned, inject 20 micro liter or as per the manufacturer’s COA and record the chromatograph.
- Calculate the tangent (theoretical plates) as per the USP method.
- The column shall be accepted for use if the values of theoretical plates obtained, exceeds the limit specified .
- If the column does not meet the requirements, the same shall be rejected & the information regarding rejection shall be given to Purchase Department.
- Record the following details of the column in the register and HPLC column usage log register.
- Column name, Dimension and serial / batch number of the column at the top of the page followed by Sr. Number, Date of analysis, Number of injection, Product, Signature of chemist and remarks.
- Allocate an in-house identification no. to the column.
- Record the requisite details of analysis on log card .
6.3 REGENERATION (During given HPLC analysis following steps should be followed)
- Flush the column, first with filtered and de-gassed water for 30 minutes at the rate of 1.0ml per minute.
- Then with either Methanol or Acetonitrile for further 30 minutes at the flow rate of 1.0ml per minute
- Condition the column with a given mobile phase for at least 30 minutes.
- Ensure that base line signal is stable, carry out the analysis.
- Check the responses of the Chromatogram of standard component with previous analytical report. If both satisfactory with respect to peak response, resolution, capacity factor, theoretical plates continue the analysis.
- If any of the above conditions are not satisfactory, regenerate the column as per the procedure given in 6.0.
- After analysis is complete flush with the mobile phase for 15 minutes (flow rate 1.0ml/min), followed by water for at least 45 minutes and finally with organic solvents Methanol or Acetonitrile.
- Extra precaution should be taken when organic buffers are being used in a mobile phase; flush the column with filtered de-gassed water for at least 1 Hour.
- Note: When inorganic buffers are used in mobile phase do not flush directly with organic solvents like Methanol, Acetonitrile as these reduces the performance & life of the column.
6.4 PROCEDURE FOR REGENERATION (REVERSED PHASE).
- Whenever Column fails in System Suitability Test as per pharmacopoeia Regeneration should be done.
- For columns C18, C8, C6, C4, Phenyl, Amino, Polymeric follow the sequence given below with a 1.0ml flow per minute flow rate.
- Flush with filtered and de-gassed water for 30 minutes. Inject 20 µl of 1.0% of Acetic acid in Water.
- Methanol for 30 minutes, inject 20µ liter of 1.0% DMSO (spectroscopic grade) in Methanol.
- Acetonitrile for 15 minutes
- Isopropanol for 15 minutes.
- Chloroform for 10 minutes.
- Acetonitrile for 10 minutes.
- Methanol for 30 minutes.
- Water for 15 minutes. Check pH of water at inlet and outlet, which should be identical.
- Mobile phase for the compound to be analyzed.
6.5 PRECAUTIONS
- Column should be kept with proper end fitting during storage.
- Do not over tighten the column end fittings.
- Frits of the column should be checked for cleanliness/disposition of the excipients in consultation with the Department Head.
- Use dedicated columns only for a given product.
- Protect the column from rapid changes in solvent’s composition.
- Do not change the flow rate at increment greater than 0.2ml/min.
- Prevent the column from mechanical shocks.
- Always use the column in the specified direction.
- pH of the mobile phase should not be less than 2.0 and not more than 7.0.
- Backpressure of the column should be monitored for variation.
- Clean the solvent reservoir filter in the mobile phase after every analysis by sonication in the Methanol for 30 minutes.
- Test sample should be clean homogeneous and free from particulate matter.
7.0 ABBREVIATIONS:
Sr. No. | Abbreviations | Details |
01. | SOP | Standard Operating Procedure |
02. | NA | Not Applicable |
03. | Sr. No. | Serial Number |
04. | QA | Quality Assurance |
05. | QC | Quality Control |
06. | DEPT. | Department |
07 | H.P.L.C. | High performance liquid chromatography |
08 | U.S.P. | United State Pharmacoepia |
09 | COA | Certificate of analysis |
10 | DMSO | Dimethyl Sulphonic Oxide |