PERFORMANCE QUALIFICATION DOCUMENT OF GC WITH HEAD SPACE ON CHROMELEON SOFTWARE
DEPARTMENT | : | QUALITY CONTROL |
LOCATION | : | INSTRUMENT ROOM |
NAME OF THE INSTRUMENT | : | GC WITH HEAD SPACE ON CHROMELEON SOFTWARE |
MAKE | : | Agilent |
MODEL/ TYPE | : | 7890B / AUTOSAMPLER & HEAD SPACE |
INSTRUMENT ID | : |
Document No. & Rev. No. | Effective Date | Reason For Revision |
New Document |
TABLE OF CONTENTS
Sr. No. | Content |
Pre-Approval | |
Objective | |
Scope | |
Qualification Team and Responsibility | |
Methodology for Performance Qualification | |
Calibration procedure | |
Calibration Observation | |
Deficiency Sheet | |
Abbreviation(s) | |
Attachment(s) | |
Summary and Conclusion | |
Post Approval |
- Pre Approval:
Prepared By:
Functional Area | Name | Designation | Signature/ Date |
Quality Control |
Reviewed By:
Functional Area | Name | Designation | Signature/Date |
Quality Assurance | |||
Head Quality Control |
Approved By:
Functional Area | Name | Designation | Signature/Date |
Head QA |
Authorized By:
Functional Area | Name | Designation | Signature/Date |
Head Quality |
- Objective:
The purpose of Performance qualification is to provide the documented evidence that the GC with Head Space on Chromeleon software is performing as per the parameters defined in performance qualification and that gives result that meet predetermined acceptance criteria.
- Scope:
This document covers all aspects of Performance qualification for GC with Head Space and Chromeleon software.
- Qualification Team and Responsibility:
- Quality Control:
- Prepares and reviews the document ensuring that the content is sufficient, clearly defined and technically accurate.
- Execution and verification of performance qualification activity as per document.
- Quality Assurance:
- Reviews the Performance qualification document.
- Ensures the proper functioning of the instrument.
- Head– QC:
- Reviews the Performance qualification document.
- Ensures the proper functioning of the equipment
- Head – QA:
- Review and approval of the Performance qualification document, the completed qualification data and the compiled report.
- Assist in the resolution of performance variances.
- Head – Quality:
- Authorization of the document for execution and after compilation.
- Methodology for Performance Qualification:
Performance qualification of GC with Chromeleon software shall be done by performing following activities.
- Calibration procedure:
- Flow Calibration
- For capillary ColumnInstall the column.Set the flow at 1.0 ml/min and check the flow at FID maintaining hydrogen, zero air and Make up flow at zero.
- Now check the flow at FID using digital flow meter.
- Set the flow at 2.0 ml/min and check the flow at FID maintaining hydrogen, zero air and Make up flow at zero. Now check the flow at FID using digital flow meter.
- Set the flow at 5.0 ml/min and check the flow at FID maintaining hydrogen, zero air and observing flow at zero.
- Now check the flow using digital flow meter.
- Set the flow at 10.0 ml/min and check the flow at FID maintaining hydrogen, zero air and observing flow at zero. Now check the flow using digital flow meter.
- Acceptance Criteria: ±5% of set value.
- Calibration of Front inlet with FID (Reproducibility and Linearity)Chromatographic Parameters & Auto sampler parameter:
Column : HP-5 (30 m x 0.32 mm ID x 0.25µm or equivalent)
Oven Temp (1) : 60°C
Initial Time (1) : 0.00 min.
Rate (1) : 25°C/min
Oven Temp (2) : 90°C
Time (2) : 0.00 min
Rate (2) : 15°C/min
Oven Temp (3) : 170°C
Final time (3) : 15.00 min
Carrier Gas : Helium/Nitrogen
Injector Temp : 250°C
Detector Temp : 300°C
Hydrogen Flow : 40 ml/min
Air Flow : 400 ml/min
Flow rate : 1.0 ml/min
Detector : FID
Make up gas : Helium/Nitrogen
Make up flow : 25 ml/min
Injection volume : 1 µl
Solvent A Pre wash : 3 Solvent B Pre wash : 3
Sample wash : 3 Sample pump : 3
Solvent A Post wash : 3 Solvent B Post wash : 3
Viscosity Delay : 0 Sec
For standard preparation refer section 5.1.3.2
- Calibration of Back inlet with FID (Reproducibility and Linearity)Chromatographic Parameters & Auto sampler parameter:
Column : HP-5 (30 m x 0.32 mm ID x 0.25µm or equivalent)
Oven Temp (1) : 60°C
Initial Time (1) : 0.00 min.
Rate (1) : 25°C/min
Oven Temp (2) : 90°C
Time (2) : 0.00 min
Rate (2) : 15°C/min
Oven Temp (3) : 170°C
Final time (3) : 15.00 min
Carrier Gas : Helium/Nitrogen
Injector Temp : 250°C
Detector Temp : 300°C
Hydrogen Flow : 40 ml/min
Air Flow : 400 ml/min
Flow rate : 1.0 ml/min
Detector : FID
Make up gas : Helium/Nitrogen
Make up flow : 25 ml/min
Injection volume : 1 µl
Split Ratio : 2:1
Solvent A Pre wash : 3 Solvent B Pre wash : 3
Sample wash : 3 Sample pump : 3
Solvent A Post wash : 3 Solvent B Post wash: 3
Viscosity Delay : 0 Sec
- Standard preparation:
Weigh about 35.0 mg of each standard of Tetradecane, Pentadecane, Hexadecane in 100.0 ml volumetric flask containing 20 ml of n-hexane and make up the volume with n-Hexane. Elution order shall be Tetradecane, Pentadecane and then Hexadecane.
- Procedure:
Inject the 1.0 µl of above standard preparation for six times and check the % area RSD and % retention time of all three components individually.
Inject 1 µl, 2 µl and 3 µl each of the above standard preparation in triplicate and calculate the correlation coefficient between injection volume and average peak area.
- Acceptance Criteria:
- RSD for peak areas of six injections should not be more than 5.0%.
- RSD for Retention time of six injections should not be more than 0.5%.
- Correlation coefficient between average area and injection volume should be more than 0.99 records the observation.
- Calibration of Head Space Sampler Reproducibility with FID DetectorChromatographic Parameters.
Column : HP-5 (30 m x 0.32 mm ID x 0.25µm or equivalent)
Oven Temp (1) : 40°C
Initial Time (T1) : 5.00 min.
Rate : 30°C/min
Oven Temp (2) : 120°C
Final Time (T2) : 5.00 min
Flow rate : 1.0 ml/min
Carrier Gas : Helium/Nitrogen
Inlet Temp : 150°C
Detector Temp : 260°C
Detector : FID
Hydrogen Flow : 40 ml/min
Air Flow : 400 ml/min
Make up flow : 20 ml/min
Split mode : 1:5
- HS Parameter
Oven Temp : 90°C
Transfer line Temp : 110°C
Vial Equil Time : 10.0 min
Injection Time : 0.5 min
Vial Shaking : OFF
Loop Temp : 100°C
GC Cycle Time : 16.0 min
Pressurization Time : 1.0 min
- Standard preparation:
Transfer 0.4 ml of ethanol in a 100.0 volumetric flask containing about 20 ml of water, shake well and makeup the volume with water.
- Procedure:
Pipette out each 2 ml of standard preparation in 6 different headspace vials, crimp it. Pipette out 2 ml of water (used for standard preparation) in two headspace vial, crimp it. (For Blank). RT of Ethanol is about 3 minute.
- Sequence:
Blank, 6 Standards and followed by Blank (For carry over).
- Acceptance Criteria:
- RSD for peak areas should not be more than 5.0%.RSD for Retention time should not be more than 0.5%.
- Carry over: NMT 0.5% .
(Area of peak in the blank after six injections x 100 / Average peak area).
- Record the observation.
- Calibration of detector linearity for FID by HeadspaceChromatographic Parameters
Column : HP-5 (30 m x 0.32 mm ID x 0.25µm or equivalent)
Oven Temp (1) : 40°C
Initial Time (T1) : 6.00 min.
Rate (R1) : 60°C/min
Oven Temp (T2) : 240°C
Final time : 8.00 min
Flow rate : 1.0 ml/min
Detector : FID
Carrier Gas : Helium/Nitrogen
Inlet Temp : 150°C
Detector Temp : 260°C
Hydrogen Flow : 40 ml/min
Air Flow : 400 ml/min
Make up flow : 20 ml/min
Split mode : 1:5
- HS Parameter
Vial Temp : 90°C
Transfer line Temp : 115°C
Vial Equil Time : 25.0 min
Pressurization Time : 5.00 min
Loop Temp : 100°C
GC Cycle Time : 30.0 min
Shaking : High (9)
Injection Time : 0.9 min
- Preparation of IPA Stock solution:
Take 0.4 ml of IPA into a 100 ml volumetric flask having about 60 ml of DMSO and then dilute it to volume with DMSO and mix well.
- Preparation of solutions for Linearity:
Linearity level | Volume of IPA stock to be taken (ml) | Dilute to 25 ml with DMSO |
1 | 0.5 | 25 |
2 | 1.0 | 25 |
3 | 2.0 | 25 |
4 | 3.0 | 25 |
5 | 4.0 | 25 |
- Pipette 2 ml of each
solution into two different vials, fit them with a septum and crimp cap and
then seal and inject. Retention time of IPA is about 3 minutes.
- Inject the each linearity level vial by headspace injection and calculate Coefficient of correlation between concentration and peak area response.
- Acceptance Criteria:
- Coefficient of correlation between Concentration and peak area is not less than 0.99 and record the observation.
- Calibration of Noise for FID
- Chromatographic Parameters
Column and Chromatographic parameters given below.
Column : 5 m x 0.1 mm uncoated fused silica
(19091 – 60620) or 19091 – 60620E)
Column flow : 0.2 ml/min
Inlet temperature : 250°C
Detector temp : 300°C
Oven temp ‘1’ : 200°C
Initial time ‘1’ : 10.5 min
Rate ‘1’ : 0.0°C/min.
Hydrogen flow : 40.0 ml/min.
Air flow : 400 ml/min.
Make up flow : 20 ml/min
Specified report : performance + Noise
After the test run print the report from system.
- Acceptance Criteria:
Initial signal ≤ 25 pA
ASTM noise ≤ 0.10 pA
Drift ≤ 2.5 / pA / hr
- Record the observation.
- Performance verification of Column OvenSet the Column oven temperature as mention below
Column Oven | Temperature to be verified | Acceptance criteria |
40°C | ± 3°C | |
100°C | ± 4°C | |
200°C | ± 5°C | |
300°C | ± 5°C |
- Turn on the oven and set the
oven temperature to 40°C and wait for at least 30 minutes after the oven
temperature goes ready for temperature stabilization.
- Take the first temperature reading from the calibrated digital thermometer and record the temperature.
- Repeat the procedure for 100°C, 200°C and 300°C also follow the above point.
- Performance verification gas flowsSet the gas flow as mention below.
Hydrogen flow | Air Flow | Make up flow | Acceptance criteria |
40 ml/min | Off | Off | 40 ± 4.0 ml/min |
Off | 400 ml/min | Off | 400 ± 40.0 ml/min |
Off | Off | 20 ml/min | 20 ± 2.0 ml/min |
- Connect the flow meter to
the detector exit.
- Record the observed value.
- An acceptance criterion is ± 10%.
- TCD DETECTOR SENSITIVITYInstall the Column HP-5 (30m x 0.32mm x 0.25µm) or equivalent.
- Enter the following parameters for calibration:
Rate (°C/min) Temp. (°C) Time
– 60 0
25 90 0
15 180 4.5
Oven equilibration time : 1.0 min
Injector temperature : 250°C
Inlet pressure : 25 psi
Flow mode : Constant Pressure
Detector temperature : 300°C
Make up flow (He) : 2 ml/min
Reference flow (He) : 20 ml/min
Polarity : Positive
Injection volume : 1.0µl
Solvent A Pre wash : 3 Solvent B Pre wash : 3
Sample wash : 3 Sample pump : 3
Solvent A Post wash : 3 Solvent B Post wash : 3
Viscosity Delay : 0 Sec
- Standard preparation:
Weigh about 35.0 mg of each standard of Tetradecane, Pentadecane, Hexadecane in 100.0 ml volumetric flask containing 20 ml of n-hexane and make up the volume with n-Hexane.
- Use n-Hexane as blank.
- Inject six injections of test mixture in the gas chromatograph and record the chromatogram.
- From the data obtained calculate the relative standard deviation of the ratio of Pentadecane to internal standard (Tetradecane) and Hexadecane to internal standard (Tetradecane) for six replicate injections separately.
- Calculate the S/N ratio for hexadecane peak through software for each injection and calculate the Average S/N ratio of six replicate injections.
- ACEPTANCE CRITERIA:
- The % RSD of the area ratio of each component i.e. Pentadecane to internal standard (Tetradecane) and Hexadecane to internal standard (Tetradecane) for six replicate injections should not be more than 8.0 %.
- The average of S/N ratio should not be less than 10 for Hexadecane peak.
- Precaution:
- Check the gas lines after connecting the column for leakages.
- If retention time of the peaks varies check the septum and if found damaged replace it.
- Set hydrogen and air flow rates properly to avoid permanent damage to the flame Ionization detector or Thermal conductivity detector (Which is applicable).
- After completion of the analysis switch of the PC, off the Hydrogen gas flow and Zero air flow.
- Bring down the temperature of injector and detector to ambient conditions.
- After attaining room temperature disconnect the column from the detector end and again increase the oven temperature to 20°C higher than the testing temperature or the column maximum temp.
- (Which is lower) and allow the carrier gas (Nitrogen) to flow through the column for a period of half an hour to flush any residuals from the column.
- Decrease the oven temperature to room temperature and then switch off the oven, switch off the carrier gas flow and power supply to the instrument.
- Remove the column from the oven and keep it at appropriate place.
- Calibration Observations:Summary
Sr. No. | Test | Acceptance Criteria | Observations | Remarks | ||||
1. | Flow Rate calibration For Capillary Column: | |||||||
1.0 ml/min | 0.95 to 1.05 ml/min | |||||||
2.0 ml/min | 1.90 to 2.10 ml/min | |||||||
5.0 ml/min | 4.75 to 5.25 ml/min | |||||||
10.0 ml/min | 9.50 to 10.50 ml/min | |||||||
2. | Calibration of Front Inlet with Flame Ionization Detector ( FID ) Reproducibility and Linearity : | |||||||
2.1 | Reproducibility of Area | |||||||
Tetradecane | % RSD of Area NMT 5.0. | |||||||
Pentadecane | % RSD of Area NMT 5.0. | |||||||
Hexadecane | % RSD of Area NMT 5.0. | |||||||
2.2 | Reproducibility of Retention Time | |||||||
Tetradecane | % RSD of Retention Time NMT 0.5. | |||||||
Pentadecane | % RSD of Retention Time NMT 0.5. | |||||||
Hexadecane | % RSD of Retention Time NMT 0.5. | |||||||
2.3 | Linearity | |||||||
Tetradecane | Correlation coefficient Should not be less than 0.99 | |||||||
Pentadecane | Correlation coefficient Should not be less than 0.99 | |||||||
Hexadecane | Correlation coefficient Should not be less than 0.99 | |||||||
3. | Calibration of Back Inlet with Flame Ionization Detector ( FID ) Reproducibility and Linearity : | |||||||
3.1 | Reproducibility of Area | |||||||
Tetradecane | % RSD of Area NMT 5.0. | |||||||
Pentadecane | % RSD of Area NMT 5.0. | |||||||
Hexadecane | % RSD of Area NMT 5.0. | |||||||
3.2 | Reproducibility of Retention Time | |||||||
Tetradecane | % RSD of Retention Time NMT 0.5. | |||||||
Pentadecane | % RSD of Retention Time NMT 0.5. | |||||||
Hexadecane | % RSD of Retention Time NMT 0.5. | |||||||
3.3 | Linearity | |||||||
Tetradecane | Correlation coefficient Should not be less than 0.99 | |||||||
Pentadecane | Correlation coefficient Should not be less than 0.99 | |||||||
Hexadecane | Correlation coefficient Should not be less than 0.99 | |||||||
4. | Calibration of Head Space Sampler Reproducibility with Flame Ionization Detector ( FID ) | |||||||
4.1 | Reproducibility of Area | |||||||
Ethanol | % RSD of Area NMT 5.0. | |||||||
4.2 | Reproducibility of Retention Time | |||||||
Ethanol | % RSD of Retention Time NMT 0.5. | |||||||
5. | Carry Over | |||||||
Ethanol | Not more than 0.5 %. | |||||||
6. | Calibration of Detector Linearity for FID by Headspace | |||||||
6.1 | Detector Linearity | Correlation coefficientShould not be less than 0.99 | ||||||
7. | Calibration of Detector Noise for Flame Ionization Detector ( FID ) | |||||||
7.1 | Noise | Initial Signal ≤ 25 pA | ||||||
ASTM Noise ≤ 0.10 pA | ||||||||
Drift ≤ 2.5 / pA / hr | ||||||||
8. | Performance Verification of Column Oven Temperature: | |||||||
8.1 | Column Oven | 40 ± 3.00C | ||||||
100 ± 4.00C | ||||||||
200 ± 5.00C | ||||||||
300 ± 5.00C | ||||||||
9. | Performance Verification of gas flow: | |||||||
Hydrogen | 40 ± 4.0 ml/min | |||||||
Air | 400 ± 40.0 ml/min | |||||||
Make Up | 20 ± 2.0 ml/min | |||||||
10. | Thermal Conductivity Detector ( TCD ) Sensitivity | |||||||
10.1 | Pentadacane to Internal standard (Tetradecane) | % RSD of the area ratio of Pentadacane to Internal standard (Tetradecane) for six replicate injections should not be more than 8.0. | ||||||
Hexadacane to Internal standard (Tetradecane) | % RSD of the area ratio of Hexadacane to Internal standard (Tetradecane) for six replicate injections should not be more than 8.0 | |||||||
10.2 | S/N Ratio | Average of S/N ratio should not be less than 10 for Hexadecane peak. | ||||||
Remarks: The observed values comply / does not comply the acceptance criteria and the equipment can be / cannot be used for routine analysis.
Done by Checked by
Date Date
- Flow Rate Calibration for Capillary column :
Flow Meter ID No. ___________ Calibration Due Date ______________
S.No. | Flow Rate ( ml per minute) | Observed Flow ( ml per minute ) | Limit ( ml per minute) |
1 | 1.0 | 0.95 to 1.05 | |
2 | 2.0 | 1.90 to 2.10 | |
3 | 5.0 | 4.75 to 5.25 | |
4 | 10.0 | 9.50 to 10.50 |
Acceptance Criteria:
Flow Rate shall be between ± 5 %.
For flow 1.0 ml/minute: 0.95 ml to 1.05 ml.
For flow 2.0 ml/minute: 1.90 ml to 2.10 ml.
For flow 5.0 ml/minute: 4.75 ml to 5.25 ml.
For flow 10.0 ml/minute: 9.50 ml to 10.50 ml.
Remarks: Complies/does not comply as per above calibration.
Calibration done by/Date: ______________ Checked By/Date: _______________
Front Inlet with Flame Ionization Detector (FID) Reproducibility and Linearity:
Balance ID No. __________________ Calibration Due Date: _________________
Batch/Lot No. Of Tetradecane ________________________
Batch/Lot No. Of Pentadecane _______________________
Batch/Lot No. Of Hexadecane ________________________
Standard Preparation:
Weigh about (35.0 mg) _________ mg of Tetradecane, (35.0 mg) ___________ mg of Pentadecane and (35.0 mg)_____________ mg of Hexadecane in (100.0 ml) ________ volumetric flask containing (20 ml) _______ of n-Hexane and make up the volume with n-Hexane.
Reproducibility of Retention Time:
S.No. | Injection No. | Retention Time of Tetradecane | Retention Time of Pentadecane | Retention Time of Hexadecane |
1 | ||||
2 | ||||
3 | ||||
4 | ||||
5 | ||||
6 | ||||
Average | ||||
% RSD |
Reproducibility of Area:
S.No. | Injection No. | Area of Tetradecane | Area of Pentadecane | Area of Hexadecane |
1 | ||||
2 | ||||
3 | ||||
4 | ||||
5 | ||||
6 | ||||
Average | ||||
% RSD |
Linearity:
Tetradecane:
S.No. | Injection Volume | Area Counts | Mean Area Counts | Correlation coefficient |
1 | 1 µl | |||
2 | 2 µl | |||
3 | 3 µl | |||
Pentadecane:
S.No. | Injection Volume | Area Counts | Mean Area Counts | Correlation coefficient |
1 | 1 µl | |||
2 | 2 µl | |||
3 | 3 µl | |||
Hexadecane:
S.No. | Injection Volume | Area Counts | Mean Area Counts | Correlation coefficient |
1 | 1 µl | |||
2 | 2 µl | |||
3 | 3 µl | |||
Acceptance Criteria:
- % RSD for Peak areas of six injections should not be more than 5.0.
- % RSD for Retention Time of six injections should not be more than 0.5.
- Correlation coefficient between average area and injection volume should not be less than 0.99.
Remarks: Complies/does not comply as per above calibration.
Calibration done by /Date: ______________ Checked By/Date: _______________
- Back Inlet with Flame Ionization Detector (FID) Reproducibility and Linearity:
Balance ID No. __________________ Calibration Due Date: _________________
Batch/Lot No. Of Tetradecane ________________________
Batch/Lot No. Of Pentadecane _______________________
Batch/Lot No. Of Hexadecane ________________________
Standard Preparation:
Weigh about (35.0 mg) _________ mg of Tetradecane, (35.0 mg) ___________ mg of Pentadecane and (35.0 mg) _____________ mg of Hexadecane in (100.0 ml) ________ volumetric flask containing (20 ml) _______ of n-Hexane and make up the volume with n-Hexane.
Reproducibility of Retention Time:
S.No. | Injection No. | Retention Time of Tetradecane | Retention Time of Pentadecane | Retention Time of Hexadecane |
1 | ||||
2 | ||||
3 | ||||
4 | ||||
5 | ||||
6 | ||||
Average | ||||
% RSD |
Reproducibility of Area:
S.No. | Injection No. | Area of Tetradecane | Area of Pentadecane | Area of Hexadecane |
1 | ||||
2 | ||||
3 | ||||
4 | ||||
5 | ||||
6 | ||||
Average | ||||
% RSD |
Linearity:
Tetradecane:
S.No. | Injection Volume | Area Counts | Mean Area Counts | Correlation coefficient |
1 | 1 µl | |||
2 | 2 µl | |||
3 | 3 µl | |||
Pentadecane:
S.No. | Injection Volume | Area Counts | Mean Area Counts | Correlation coefficient |
1 | 1 µl | |||
2 | 2 µl | |||
3 | 3 µl | |||
Hexadecane:
S.No. | Injection Volume | Area Counts | Mean Area Counts | Correlation coefficient |
1 | 1 µl | |||
2 | 2 µl | |||
3 | 3 µl | |||
Acceptance Criteria:
- % RSD for Peak areas of six injections should not be more than 5.0.
- % RSD for Retention Time of six injections should not be more than 0.5.
- Correlation coefficient between average area and injection volume should not be less than 0.99.
Remarks: Complies/does not comply as per above calibration.
Calibration done by /Date: ______________ Checked By/Date: _______________
- Calibration of Head Space Sampler Reproducibility with Flame Ionization Detector ( FID ):
Batch/Lot No. Of Ethanol ________________________
Standard Preparation:
Transfer (0.4 ml) _______ of ethanol in a (100.0ml) _________ volumetric flask containing about 20 ml of water shake well and make up the volume with water.
S.No. | Vial No. | Retention Time of Ethanol Peak | Area of Ethanol Peak |
1 | |||
2 | |||
3 | |||
4 | |||
5 | |||
6 | |||
Average | |||
% RSD NMT 5.0 % for Area and NMT 0.5 % for Retention Time |
- Carry Over:
% Carry Over = Area of peak in the blank after six injection
______________________________________ x 100
Average Peak area of Standard (Ethanol)
S.No. | Test | Observation |
1 | Area of Ethanol peak in Blank | |
2 | Average Area of Ethanol | |
3 | % Carry over |
Acceptance Criteria:
- % RSD for Peak areas of six standards should not be more than 5.0.
- % RSD for Retention Time of six standards should not be more than 0.5.
- Carry Over not more than 0.5 %.
Remarks: Complies/does not comply as per above calibration.
Calibration done by /Date: ______________ Checked By/Date: _______________
- Calibration of Detector Linearity for FID by Headspace:
Batch/Lot No. of IPA _____________
Preparation of IPA Stock Solution:
Take (0.4ml) ______ ml of IPA into a (100.0 ml) ______ ml volumetric flask having about (60 ml) ______ of DMSO and then dilute it to volume with DMSO and mix well.
Preparation of Linearity Solution Level 1:
Dilute (0.5ml) ______ ml of IPA stock solution to (25.0ml) _________ ml with DMSO.
Preparation of Linearity Solution Level 2:
Dilute (1.0ml) ______ ml of IPA stock solution to (25.0ml) _________ ml with DMSO.
Preparation of Linearity Solution Level 3:
Dilute (2.0ml) ______ ml of IPA stock solution to (25.0ml) _________ ml with DMSO.
Preparation of Linearity Solution Level 4:
Dilute (3.0ml) ______ ml of IPA stock solution to (25.0ml) _________ ml with DMSO.
Preparation of Linearity Solution Level 5:
Dilute (4.0ml) ______ ml of IPA stock solution to (25.0ml) _________ ml with DMSO.
Linearity Level | Area Counts | Concentration | Correlation coefficient |
1 | |||
2 | |||
3 | |||
4 | |||
5 |
Acceptance Criteria:
Correlation of coefficient between concentration and peak area is not less than 0.99.
Remarks: Complies/does not comply as per above calibration.
Calibration done by /Date: ______________ Checked By/Date: _______________
- Calibration of Detector Noise for Flame Ionization Detector ( FID ):
S.No. | Test | Observation |
1 | Initial Signal ≤ 25 pA | |
2 | ASTM Noise ≤ 0.10 Pa | |
3 | Drift ≤ 2.5 / pA / hr |
Acceptance Criteria:
- Initial Signal ≤ 25 pA
- ASTM Noise ≤ 0.10 pA
- Drift ≤ 2.5 / pA / hr
- ..
Remarks: Complies/does not comply as per above calibration.
Calibration done by /Date: ______________ Checked By/Date: _______________
- Performance Verification of Column Oven Temperature:
Digital Thermometer ID No. ______________ Calibration Due Date ___________________
S.No. | Set Temperature ( ˚C) | Observed Temperature ( ˚C) | Limit ( ˚C) |
1 | 40 | 37.0 to 43.0 | |
2 | 100 | 96.0 to 104.0 | |
3 | 200 | 195.0 to 205.0 | |
4 | 300 | 295.0 to 305.0 |
Acceptance Criteria:
- For Set Column Oven Temperature 40˚C ± 3˚C.
- For Set Column Oven Temperature 100˚C ± 4˚C.
- For Set Column Oven Temperature 200˚C ± 5˚C.
- For Set Column Oven Temperature 300˚C ± 5˚C.
Remarks: Complies/does not comply as per above calibration.
Calibration done by /Date: ______________ Checked By/Date: _______________
- Performance Verification of gas flow:
Flow Meter ID No. __________________ Calibration Due Date: _______________
S.No. | Gas Type | Flow Rate ( ml per minute) | Observed Flow ( ml per minute) | Limit ( ml per minute) |
1 | Hydrogen | 40 | 36 to 44 | |
2 | Air | 400 | 360 to 440 | |
3 | Make up | 20 | 18 to 22 |
Acceptance Criteria:
- For Gas type Flow Rate for Hydrogen 40 ± 4.0 ml/min.
- For Gas type Flow Rate for Flow Air 400 ± 40.0 ml/min.
- For Gas type Flow Rate for Flow make up 20 ± 2.0 ml/min.
Remarks: Complies/does not comply as per above calibration.
Calibration done by /Date: ______________ Checked By/Date: _______________
- Thermal Conductivity Detector (TCD) Sensitivity:
Balance ID No. __________________ Calibration Due Date: _________________
Batch/Lot No. Of Tetradecane ________________________
Batch/Lot No. Of Pentadecane _______________________
Batch/Lot No. Of Hexadecane ________________________
Standard Preparation:
Weigh about (35.0 mg) _________ mg of Tetradecane, (35.0mg) ___________ mg of Pentadecane and (35.0mg) _____________ mg of Hexadecane in (100.0 ml) ________ volumetric flask containing (20 ml) _______ of n-Hexane and make up the volume with n-Hexane.
% RSD of the area ratio of Pentadecane to Internal standard (Tetradecane):
Sr. No. | Injection No. | Area of Pentadecane | Area of Tetradecane | Area ratio of Pentadecane to Tetradecane |
1 | ||||
2 | ||||
3 | ||||
4 | ||||
5 | ||||
6 | ||||
Average of Area Ratio | ||||
% RSD of Area Ratio |
% RSD of the area ratio of Hexadecane to Internal standard (Tetradecane):
Sr.No. | Injection No. | Area of Hexadecane | Area of Tetradecane | Area ratio of Hexadecane to Tetradecane |
1 | ||||
2 | ||||
3 | ||||
4 | ||||
5 | ||||
6 | ||||
Average of Area Ratio | ||||
% RSD of Area Ratio |
S/N ratio of Hexadecane:
Sr. No. | Injection No. | S/N ratio of Hexadecane |
1 | ||
2 | ||
3 | ||
4 | ||
5 | ||
6 | ||
Average |
Acceptance Criteria:
- The % RSD of the area ratio of each component i.e. Pentadecane to internal standard (Tetradecane) and Hexadecane to internal standard (Tetradecane) for six replicate injections should not be more than 8.0.
- The average of S/N ratio should not be less than 10 for Hexadecane peak.
Remarks: Complies/does not comply as per above calibration.
Calibration done by /Date: ______________ Checked By/Date: _____________
- DEFICIENCY SHEET:
Report any deficiencies from the acceptance criteria or from protocol instructions in the ‘Deficiency report form’ attached as Annexure I. Record the Deficiency report number and brief description in the table below. Include all deficiency report forms as Annexure I. Indicate the completeness checked by sign and date after resolving of the deficiency.
Sr. No. | Deviation Report No. | Brief Description | Checked for completeness By Sign./Date |
- ABBREVIATION(S):
SOP : Standard Operating Procedure
QA : Quality Assurance
HOD : Head of department
QC : Quality Control
GC : Gas Chromatograph
Min : Minute
RSD : Relative Standard Deviation
RT : Retention Time
NLT : Not Less Than
NMT : Not More Than
FID : Flame Ionization Detector
TCD : Thermal Conductivity Detector
- Attachment(S) :
Sr. No. | Annexure No. | Document Title |
Remarks:
______________________________________________________________________________________________________________________________________________________________________
Verified By: ________________ (Sign./Date)
- Summary and conclusion:
___________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________________
Prepared By: ___________________________________________ __________________
Name / Department Sign. / Date
- Post Approval:
Executed By:
Functional Area | Name | Designation | Signature/ Date |
Quality Control |
Reviewed By:
Functional Area | Name | Designation | Signature/Date |
Quality Assurance | |||
Head Quality Control |
Approved By:
Functional Area | Name | Designation | Signature/Date |
Head QA |
Authorized By:
Functional Area | Name | Designation | Signature/Date |
Head Quality |