CALIBRATION RECORD OF FTIR

Procedure: The wave-number scale may be verified using a calibrated polystyrene film, which has transmission minima (absorption maxima) at the wave numbers (in cm–1) shown in below:

S. No. Standard wave number ( cm-1) Observation ( cm-1) Tolerance
01 3060.0   ± 1.0 cm-1
02 2849.5   ± 1.0 cm-1
03 1942.9   ± 1.0 cm-1
04 1601.2   ± 1.0 cm-1
05 1583.0   ± 1.0 cm-1
06 1154.5   ± 1.0 cm-1
07 1028.3   ± 1.0 cm-1

Result: Pass /Fail

Calibration Status                 :                              

Corrective action (If any)    :