CALIBRATION RECORD OF REFRACTOMETER

Reference Liquid Room Temp. 25.00C ± 20C Observed Refractive Index Limit Remark
Water     1.3320 – 1.3330  
Carbon Tetrachloride     1.4600 – 1.4610  
Toluene     1.4940 – 1.4970  

Result: Pass /Fail

Calibration Status                  :                              

Corrective action (If any)      :