- Objective
To lay down the procedure for Cleaning, Operation and calibration of Tablet Friability apparatus.
- Scope
- Responsibility
QC chemist / Officer : To follow the procedure
- Accountability
Head Quality Control : For SOP compliance
- Procedure
- Cleaning
- Before cleaning switch off the main.
- Clean the surface of instrument by using dry lint free cloth.
- Press the lock button remove out from the shaft then remove the drum (D 2) and open
- Cleaning
the lid and remove the dust powder available inside of the drum and then thoroughly clean with wet lint free cloth and followed by dry lint free cloth.
- Repeat the same above procedure for drum (D1) cleaning.
- Slide out the tray from their rack and clean first with wet lint free cloth and followed
by dry lint free cloth.
- Operation:
- Ensure the cleaning and calibration of the instrument.
- Check the drum is fixed and locked properly with shaft and trays are arranged below the drum.
- Tilt the instrument (10°) if required.
- Switch ON the instrument, drum will initialize itself to the loading position, the display will now show START.
- Weigh and record the weight of the tablets.
- Adjust the counts to 100 by pressing the COUNT key followed by ‘1’, ‘0’, ‘0’ keys. (Count range 1 to 99999).Press ENTER to confirm the reading.
- To see and confirm the number of counts press COUNT key .
- On confirming the number of counts press RUN/HALT key to start, the display shows the elapsed count.
- Test can be performed by adjusting ‘TIME’ similarly as ‘COUNT’.
- When test is over drum rotates in reverse direction discharging the tablets in the tray.
- The test over is indicated by an audible beep and display shows END.
- The drum initializes itself to loading position and display shows START indicating the instrument is ready for the next run.
- Remove the tablets from the tray.
- De- dust and weigh the tablets and note down the weight of the tablets
- Method:
- Sample Quantity: For tablets with an average weight of 0.650 gm or less take a sample of whole tablets to about 6.50 gm and for tablets with an average weight of more than 0.650 gm take a sample of 10 whole tablets. .
- De-dust the tablets carefully.
- Weigh accurately the required number of tablets and record in gram (A).
- Place the tablets in the drum and operate the apparatus as per SOP for 100 rounds.
- After completion of test, remove the tablets, remove any dust from them and weigh them
- Accurately and record in gram (B) and calculate as per the formula.
- Calculation:
- Method:
[A] – [B]
Friability = ———————x 100
[A]
- Acceptance Criteria:
- Not more than 1 % w/w.
- If cracked, cleaved or broken tablets are present in the tablet sample after test, the
sample fails.
- If the weight loss is more than 1 % w/w, repeat the test twice and the mean of (3 tests) determined. This should not be more than 1 %.
- Calibration procedure
- Ensure that the equipment shall be clean.
- Operate the apparatus as per the operation point No.5.2
- Affix metal piece on the drum and check that the tachometer will sense it.
- Run the apparatus and wait until reading on display of tachometer stabilizes.
- After stabilization record the observation on annexure I.
- Repeat the above process twice. And calibrate the timer with the help of stop watch.
- Acceptance Criteria: 25 ± 1 RPM (24-26) and time ( 00:04:00±10sec).
- Frequency: Monthly ± 3 Days
- If the variation is not with in the specified tolerance limit; affix the status as out of calibration and inform to department head.
- Do not use the equipment until all error are rectified, after rectification again calibrate the equipment and record the observation.
- Calibration procedure
- Annexure
Annexure I: Calibration Record of friability test apparatus.
- Reference (S)
IP / USP
- Glossary
SOP | : | Standard Operating Procedure |
BMR | : | Batch Manufacturing Record |
NA | : | Not Applicable |
RPM | : | Rotation per minute |
% | : | Percentage |
- Revision History
Rev. No. | Details of changes | Reason for change |