OPERATION AND CALIBRATION OF COLONY COUNTER

1.0 OBJECTIVE:

This SOP describes the procedure for operation and calibration of colony counter.

2.0 SCOPE:

This SOP is applicable to the procedure for operation and calibration of colony counter.

3.0 RESPONSIBILITY:

         Microbiologist

4.0 ACCOUNTABILITY:

          QC Manager.

5.0 PROCEDURE:

5.1 Operating Procedure:

  • Ensure that the instrument is clean and suitable for use.
  • Ensure that the magnifying lens is fixed on its appropriate plate.
  • Ensure that the marker is connected to either right or left side of the plate.
  • Switch on the mains.
  • Switch on the power button of the instrument. The internal light will be flashed “ON” & the digital screen will show “0000” figure in red.
  • Now press the plate containing colonies to be counted.
  • Now, mark each & every colony one after by other by pressing the tip of the marker.
  • The number of colonies will automatically get display on the screen can be set to”0000” by pressing of marker.
  • If any correction in current counting of colony; correct the count by Count Correction key.
  • During the counting of colony, hold the counting by press the Hold key.
  • After completion of counting the colony, for next colony counting press the Reset Key.
  • After completion the work, switch ‘OFF’ the Instrument.
  • Maintain the usage record of colony counter .

5.2 Calibration:

  • Operate the instrument as per the SOP & press the tip of the marker 10 times, 50 times & 100 times one by one then check the number displayed on the screen. This Method applied minimum three times after Reset the instruments. And record maintain .
  • Frequency: Every Month.

6.0 TRAINING:

Trainer   : Manager – Quality Control

Trainees : Staff of the microbiology departments

7.0 DISTRIBUTION:

Master Copy                  :           QA Department

Controlled Copy            :           Microbiology Department

Display Copy                 :           Microbiology Department (If Required)

8.0 ATTACHMENTS:

Sr. No. Title
01. Usage Record of Colony Counter
02. Calibration Record of Colony Counter

9.0 REFERENCES:

In-House

10.0 ABBREVIATIONS:

Abbreviations Extended Form
SOP Standard Operating Procedure
NA Not Applicable
MB Microbiology
Sr. No. Serial Number
QA Quality Assurance
DEPT. Department
IPA Isopropyl Alcohol
% Percentage

11.0 REVISION HISTORY OF CHANGE:

Sr. No. Date Revision Details Revision No.
01 ……….. NEW SOP 00

Bhanu Pratap Singh