SOP for operation & calibration of Disintegration Test Apparatus

  1. OBJECTIVE:

To lay down a procedure for Operation & Calibration of Disintegration test apparatus.

  • SCOPE:

This procedure is applicable for procedure for Operation & Calibration of Disintegration test apparatus.

  • RESPONSIBILITY:

QC Officer/ QC Executive

  • ACCOUNTABILITY:

QC Manager.

  • PROCEDURE:
    • Precautions:
      • Do not switch on the mains if water in the tank is not up to the mark.
      • While placing and removing the basket assembly do not apply excessive force.
      • Do not use abrasive materials to clean the instrument.
      • Do not bend the heater while clean the water bath.
    • Operating Procedure
      • Ensure that instrument is clean and free from dust.
      • Ensure that clean purified water / DM Water is filled in the basket till the desired level.
      • Switch ON the main supply and the instrument ON/OFF switch on the instrument.
      • Switch on the heater by pressing “TEMP” Key.
      • Set the temp. by aero key at 370C.
      • Set the time by aero key the desired test t0.
      • ime.
      • Place the capsules to be tested in each of the six cylinders of basket rack assembly after achieved the desired test temperature.
      • Place the disks if mention in the monograph.
      • Press the on key the instrument.
      • Watch the complete disintegration of the capsules and note down in usage log book of disintegration test apparatus.
      • Press stop key to stop the test.
    • Calibration:
  • Switch on the mains and switch on the instrument.
  • Ensure that the beakers are filled with water the DM water.
  • Press the Temp. key for set the temperature at 370C and press enter key.
  • Allow the temperature to reach to 370C.When the display shows the temperature 370C, insert the calibrated thermometer into the beaker.
  • Note down the reading in calibration record of disintegration test apparatus instrument operate for 2 hours (120 min) and check the temperature after 10 minute with calibrated thermometer and note down the reading.
  • Repeat the two more time to confirm the result.
  • Results are found satisfactory if the temperature lies between 370C ± 20C.
  • For Cycles per minute: Note down the number of cycles in one minute; note down the reading after 10 minutes for 2 hour (120 minute).
  • Repeat the test two more times to confirm the results.
  • Result are found satisfactory if the cycles per minute 29 to 32.
  • For Distance Travelling: Measure the distance travel of basket by calibrated scale.
  • Result are found satisfactory if the distance travelling is between 53 mm to 57 mm.Frequency  – one in month.
  • TRAINING:

Trainer   : Manager – Quality Control

Trainees : Staff of the QC departments

  • DISTRIBUTION:

Master Copy                  :           QA Department

Controlled Copy            :           QC Department

Display Copy                 :           QC Department

  • ATTACHMENTS:
Sr. No. Title Annexure No.
01. NA NA
  • REFERENCES:

In-House

  1. ABBREVIATIONS:
Abbreviations Extended Form
SOP Standard Operating Procedure
NA Not Applicable
QC Quality Control
Sr. No. Serial Number
QA Quality Assurance
DEPT. Department
  1. REVISION HISTORY OF CHANGE:
Sr. No. Date Revision Details Revision No.