SOP for operation & calibration of Hot Plate

  1. OBJECTIVE:

To lay down a procedure for Operation & Calibration of hot plate.

  • SCOPE:

This procedure is applicable for procedure for Operation & Calibration of hot plate.

  • RESPONSIBILITY:

QC Officer/ QC Executive

  • ACCOUNTABILITY:

QC Manager.

  • PROCEDURE:
    • Operation Procedure:
      • Ensure that the instrument is clean and suitable for use if not, than clean the all parts of the instrument with clean & dry cloth or by tissue paper.
      • Switch ‘ON’ the main power supply of the instrument.
      • Set the required temp. Speed low, medium & high with the help of knob.
      • Put the sample container on hot surface.
      • After completion of the work switch ‘OFF’ the main power supply.
    • Calibration:
      • Operate the instrument according to the operating instructions.
      • Take the 100 ml of purified water in 500ml of glass beaker; set the temperature at 50, 60, 70, 80, 90, 1000C and record the results in Calibration Record.
      • Limit of temperature variation is ± 2ºC from the desired temperature.
      • Frequency of oven calibration is once in three months.
    • Precautions:
      • Instrument should be handled properly.
      • If the instrument does not produce satisfactory results it should be labeled ‘Under Maintenance’ or ‘Out of Order’. 
    • Frequency:
    • Calibration Frequency: Every month.
  • TRAINING:

Trainer   : Manager – Quality Control

Trainees : Staff of the QC departments

  • DISTRIBUTION:

Master Copy                  :           QA Department

Controlled Copy            :           QC Department

Display Copy                 :           QC Department

  • ATTACHMENTS:

NA

  • REFERENCES:

In-House

  1. ABBREVIATIONS:
Abbreviations Extended Form
SOP Standard Operating Procedure
NA Not Applicable
QC Quality Control
Sr. No. Serial Number
QA Quality Assurance
DEPT. Department
0C Centigrade
  1. REVISION HISTORY OF CHANGE:
Sr. No. Date Revision Details Revision No.
01 15/01/2020 NEW SOP 00

Bhanu Pratap Singh

BHANU PRATAP SINGH IS EXPERIENCED IN PHARMACEUTICAL, AUTHOR AND FOUNDER OF PHARMACEUTICAL GUIDESLINE (WWW.PHARMAGUIDESLINE.COM), A WIDELY READ PHARMACEUTICAL BLOG SINCE 2019. EMAIL:- INFO@PHARMAGUIDESLINE.COM

View all posts by Bhanu Pratap Singh →

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!