SOP for operation & calibration of Ultra Sonic Bath

  1. OBJECTIVE:

To lay down a procedure for Operation & Calibration of ultrasonic bath.

  • SCOPE:

This procedure is applicable for procedure for Operation & Calibration of ultrasonic bath.

  • RESPONSIBILITY:

QC Officer/ QC Executive

  • ACCOUNTABILITY:

QC Manager.

  • PROCEDURE:
    • Operating Procedure
      • Ensure that the instrument is clean and suitable for use.
      • Ensure that all the switches are ‘OFF’ in normal position and the level of water in the ultrasonic is up to the basket.
      • Switch ‘ON’ the main power supply of the instrument.
      • Set the required time for sonication with push the CHK button by navigation key.
      • Put the sample container into the basket of ultrasonic
      • Press the ‘ON’/OFF button which is located on the front body of the ultrasonic.
      • After completion of the time ultrasonic stop automatically.
      • After completion of the work switch ‘OFF’ the main power supply.
      • Change the water, when it seems dirty or turbid.
      • Maintain its usage log book.
  • Calibration:
  • Calibrate the ultrasonic Timer by using the calibrated stopwatch.
  • Switch ‘ON’ the main power supply unit and set the time 5 min, 10 min, and 15min respectively and match the time with Calibrated stop watch.
  • Maintain the calibration record of ultrasonic Timer.
  • Frequency of the calibration of ultrasonic Timer is once in 3 months.
    • Precautions:
      • Instrument should be handled properly.
      • If the instrument does not produce satisfactory results it should be labeled ‘Under Maintenance’ or ‘Out of Order’.
  • TRAINING:

Trainer   : Manager – Quality Control

Trainees : Staff of the QC departments

  • DISTRIBUTION:

Master Copy                  :           QA Department

Controlled Copy            :           QC Department

Display Copy                 :           QC Department (If Required)

  • ATTACHMENTS:

NA

  • REFERENCES:

In-House

  1. ABBREVIATIONS:
Abbreviations Extended Form
SOP Standard Operating Procedure
NA Not Applicable
QC Quality Control
Sr. No. Serial Number
QA Quality Assurance
DEPT. Department
  1. REVISION HISTORY OF CHANGE:
Sr. No. Date Revision Details Revision No.
01 02/02/2017 NEW SOP 00

Bhanu Pratap Singh