Operation and calibration of Moisture Meter

  1. Objective

To lay down a procedure for Operation and calibration of Moisture Meter.

  • Scope

This SOP is applicable for Operation and calibration of Moisture Meter in Quality Control.  

  • Responsibility

QC Chemist    : Follow the procedure and maintain record as applicable.

  • Accountability

Head Quality Control  : System Compliance

  • Procedure
    • Operation:      
      • Set rotator knob to position 2.
      • Insert pins in the box in level position without shaking.
      • Insert pins in the box at three/ four places at different spots.
      • The moisture contents reading display will show on the screen.
      • Measuring range of moisture meter is 0% to 90%.
      • Enter the details in Instrument log.
    • Calibration
      • Place the pins of the Moisture meter across the two screws on the Moisture Calibrator (moisture calibrator shall be calibrated from outside Agency).
      • The knob should be kept in position 2.
      • Reading should be 12.0%±0.5%.
      • Fill the details in Annexure-I.
      • If there is any deviation in the reading, then adjust the instrument by removing the 2 screws provided on the display and then remove the cover , you can find a small calibration screw .
      • Adjust this Trimpot, so that the display reads 12.0%±0.5%.
      • Prepare the calibration tag and affix it on the equipment.
      • Attach the old calibration tag with the old calibration record and do not throw it.
      • Frequency: Yearly ±30 days.
    • Precautions:
      • Do not apply the extra force above the capacity. Otherwise the meter may be damaged.
      • Do not hit or put weight on the LED.
      • Do not press button with pointed tools, nails or speculated tools.
      • Clear the dirt and dust with soft cloth. Do not use volatile liquid to clean the meter. If there is something wrong with it, contact with service engineer.
      • After use place the moisture meter in the screw type box.
    • If the variation is not within the specified tolerance limits:
      •  Label the equipment with a tag, “OUT OF CALIBRATION”.
      •  Inform the Department Head.
  • Annexure (S)

            Annexure-I      : Calibration record of Moisture Meter.

  • Reference (S)
  •  NA
  • Glossary

SOP                            :  Standard Operating Procedure

QC                              :  Quality Control

QA                               :  Quality Assurance

No.                              :  Number

%                                 :  Percentage

  • Revision History
Rev. No. Details of changes Reason for change

Bhanu Pratap Singh

BHANU PRATAP SINGH IS EXPERIENCED IN PHARMACEUTICAL, AUTHOR AND FOUNDER OF PHARMACEUTICAL GUIDESLINE (WWW.PHARMAGUIDESLINE.COM), A WIDELY READ PHARMACEUTICAL BLOG SINCE 2019. EMAIL:- INFO@PHARMAGUIDESLINE.COM

View all posts by Bhanu Pratap Singh →

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!