- OBJECTIVE:
To lay down a procedure for Operation & Calibration of hot air oven.
- SCOPE:
This procedure is applicable for procedure for Operation & Calibration of hot air oven.
- RESPONSIBILITY:
QC Officer/ QC Executive
- ACCOUNTABILITY:
QC Manager.
- PROCEDURE:
- Operation Procedure:
- Connect the 3 pin cord to main line and switch ‘ON’ the unit, and adjust the Temp. Speed adjusting Knob on Low or medium or High as per test requirements.
- Press ‘SET’ button, the controller will show the set temperature. Change if required with the help of Coarse and Fine other two knobs while keeping ‘SET’ button pressed.
- Now the digital display indicator indicates the set temperature. If ‘SET’ button is released it will show the actual temperature of the chamber.
- To check the set temperature, press ‘SET’ button and digital display indicator show the set temperature.
- Always keep the oven Fan switch ‘ON’.
- Calibration:
- Operate the instrument according to the operating instructions 5.1.1 to 5.1.6.
- Set the temperature at 50, 60, 70, 80, 90, 100 & 150ºC and record the results in Calibration Record.
- Limit of temperature variation is ± 2ºC from the desired temperature.
- Frequency of oven calibration is once in three months.
- Operation Procedure:
- TRAINING:
Trainer : Manager – Quality Control
Trainees : Staff of the QC departments
- DISTRIBUTION:
Master Copy : QA Department
Controlled Copy : QC Department
Display Copy : QC Department
- ATTACHMENTS:
NA
- REFERENCES:
In-House
- ABBREVIATIONS:
Abbreviations | Extended Form |
SOP | Standard Operating Procedure |
NA | Not Applicable |
QC | Quality Control |
Sr. No. | Serial Number |
QA | Quality Assurance |
DEPT. | Department |
0C | Centigrade |
- REVISION HISTORY OF CHANGE:
Sr. No. | Date | Revision Details | Revision No. |
01 | 15/01/2020 | NEW SOP | 00 |